A multicenter evaluation of a novel microfluidic rapid AST assay for Gram-negative bloodstream infections
Institute of Medical Microbiology, Virology and Hygiene, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Gradientech AB, Uppsala, Sweden.
Gradientech AB, Uppsala, Sweden.
Gradientech AB, Uppsala, Sweden.
Show others and affiliations
2024 (English)In: Journal of Clinical Microbiology, ISSN 0095-1137, E-ISSN 1098-660X, Vol. 62, no 10, article id e0045824Article in journal (Refereed) Published
Abstract [en]

Common phenotypic methods for antimicrobial susceptibility testing (AST) of bacteria are slow, labor intensive, and display considerable technical variability. The QuickMIC system provides rapid AST using a microfluidic linear gradient. Here, we evaluate the performance of QuickMIC at four different laboratories with regard to speed, precision, accuracy, and reproducibility in comparison to broth microdilution (BMD). Spiked (n = 411) and clinical blood cultures (n = 148) were tested with the QuickMIC Gram-negative panel and compared with BMD for the 12 on-panel antibiotics, and 10 defined strains were run at each site to measure reproducibility. Logistic and linear regression analysis was applied to explore factors affecting assay performance. The overall essential agreement and categorical agreement between QuickMIC and BMD were 95.6% and 96.0%, respectively. Very major error, major error, and minor error rates were 1.0%, 0.6%, and 2.4%, respectively. Inter-laboratory reproducibility between the sites was high at 98.9% using the acceptable standard of ±1 twofold dilution. The mean in-instrument analysis time overall was 3 h 13 min (SD: 29 min). Regression analysis indicated that QuickMIC is robust with regard to initial inoculum and delay time after blood culture positivity. We conclude that QuickMIC can be used to rapidly measure MIC directly from blood cultures in clinical settings with high reproducibility, precision, and accuracy. The microfluidics-generated linear gradient ensures high reproducibility between laboratories, thus allowing a high level of trust in MIC values from single testing, at the cost of reduced measurement range compared to BMD.

IMPORTANCE: Increasing antimicrobial resistance underscores the need for new diagnostic solutions to guide therapy, but traditional antimicrobial susceptibility testing (AST) is often inadequate in time-critical diseases such as sepsis. This work presents a novel and rapid AST system with a rapid turnaround of results, which may help reduce the time to guided therapy, possibly allowing early de-escalation of broad-spectrum empirical therapy as well as rapid adjustments to treatments when coverage is lacking.

Place, publisher, year, edition, pages
American Society for Microbiology, 2024. Vol. 62, no 10, article id e0045824
Keywords [en]
diagnostics, multicenter study, rapid AST, sepsis
National Category
Infectious Medicine
Identifiers
URN: urn:nbn:se:oru:diva-116391DOI: 10.1128/jcm.00458-24ISI: 001319801200002PubMedID: 39324811Scopus ID: 2-s2.0-85206596482OAI: oai:DiVA.org:oru-116391DiVA, id: diva2:1901769
Funder
Swedish Agency for Economic and Regional GrowthAvailable from: 2024-09-30 Created: 2024-09-30 Last updated: 2025-04-01Bibliographically approved

Open Access in DiVA

No full text in DiVA

Other links

Publisher's full textPubMedScopus

Authority records

Sundqvist, Martin

Search in DiVA

By author/editor
Sundqvist, Martin
By organisation
School of Medical Sciences
In the same journal
Journal of Clinical Microbiology
Infectious Medicine

Search outside of DiVA

GoogleGoogle Scholar

doi
pubmed
urn-nbn

Altmetric score

doi
pubmed
urn-nbn
Total: 44 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf