Calibration and Uncertainty Estimation for Water Content Measurement in SolidsShow others and affiliations
2021 (English)In: International journal of thermophysics, ISSN 0195-928X, E-ISSN 1572-9567, Vol. 42, no 3, article id 42Article in journal (Refereed) Published
Abstract [en]
In the field of water content measurement, the calibration of coulometric methods (e.g., coulometric Karl Fischer titration or evolved water vapor analysis) is often overlooked. However, as coulometric water content measurement methods are used to calibrate secondary methods, their results must be obtained with the highest degree of confidence. The utility of calibrating such instruments has been recently demonstrated. Both single and multiple point calibration methods have been suggested. This work compares these calibration methods for the evolved water vapor analysis technique. Two uncertainty estimation approaches (Kragten's spreadsheet and M-CARE software tool) were compared as well, both based on the ISO GUM method.
Place, publisher, year, edition, pages
Springer , 2021. Vol. 42, no 3, article id 42
Keywords [en]
Calibration, Evolved water vapor analysis, Metrology in chemistry, Moisture in solids, Uncertainty estimation
National Category
Analytical Chemistry
Identifiers
URN: urn:nbn:se:oru:diva-90070DOI: 10.1007/s10765-021-02796-yISI: 000616620700002Scopus ID: 2-s2.0-85100255194OAI: oai:DiVA.org:oru-90070DiVA, id: diva2:1533057
Note
Funding Agency: