Background: The increasing use of computed tomography (CT) raises concerns about radiation-induced cancer. Simultaneously, advancements such as automatic exposure control (AEC), tube current modulation (TCM) and bowtie filter aim to reduce radiation dose without compromising image quality. Recently, 3D camera-based systems have been introduced to improved patient centering and support dose optimization. Aim: The aim of this study was to compare the effectiveness of 3D camera-based and manual centering by radiographers, of phantoms, to evaluate mis-centering errors in CT-imaging. Method: Two thoracic phantoms were scanned on two Siemens CT systems: Somatom Drive and Naeotom Alpha, at different vertical table positions from isocenter. One phantom was also scanned using both manual and 3D camera-based centering. CTDIvol values were collected for radiation dose analysis, and image noise was assessed using region of interest (ROI) standard deviation. Scans were repeated to ensure consistency, and statistical analysis was performed using the Wilcoxon signed rank test. Results: Positioning below the patient isocenter increased radiation dose, while positioning above patient isocenter reduced dose by up to 32 %. Manual centering achieved minimal deviation from isocenter and maintained lower CTDIvol. In contrast, 3D camera-based centering consistently overestimated table height, resulting in significantly increased dose values across both scanners. Discussion: Vertical miscentering affected radiation dose more than image quality. Although 3D camera-based centering was more reproducible, it was less accurate than manual positioning. Conclusion: Manual centering achieved higher accuracy than 3D camera systems. Combining skilled radiographers with automated tools offers the best strategy to ensure optimal image quality, dose efficiency and patient safety in CT imaging