2015 (English)In: Cardiovascular Electrophysiology, ISSN 1045-3873, E-ISSN 1540-8167, Vol. 26, no 10, p. 1088-1094Article in journal (Refereed) Published
Abstract [en]
Inappropriate ICD Shocks and Complications in HCM IntroductionThe expanded use of implantable cardioverter-defibrillators (ICDs) to prevent sudden cardiac death (SCD) in hypertrophic cardiomyopathy (HCM) based on risk stratification in individuals without known previous ventricular arrhythmia is justified by an acceptable risk of device-related adverse events. Such complications, leading to surgical procedures or inappropriate electrical shocks, may impact mortality, morbidity, quality of life, and cost-effectiveness. Methods and ResultsFrom the Swedish ICD Registry, implants due to HCM since 1995 until November 2012 in patients aged 18 years were identified and medical records reviewed. Inappropriate ICD shock occurred in 14.3% (46 of 321 patients; mean follow-up 5.4 years) with a recurrent episode in 28.2% of them. In multivariable analysis, hazard ratio (HR) for atrial fibrillation was 3.5 (95% confidence interval 1.8-6.8; P < 0.001) but showed no significant association to male sex (HR = 0.77), age (HR = 0.99), secondary indication (HR = 1.02) or device, ICD-DR/CRTD vs. ICD-VR (HR 1.07). Inappropriate shocks were triggered by atrial fibrillation/flutter or ectopic tachycardia (56.5%), sinus tachycardia (14.5%), lead dysfunction (14.5%), and T-wave oversensing (13.0%). A reintervention, besides elective device replacement, occurred in 92 patients (totally 150 procedures). The majority were lead-related (70.0%) procedures, especially of the ICD lead. Reintervention was associated with female sex (HR = 1.6 P = 0.04). ConclusionInappropriate ICD shock triggered by atrial arrhythmias, lead dysfunction, or complications requiring surgical interventions, is a concern in HCM patients who will be eligible for long-term prevention of sudden death. Efforts to avoid adverse events and provide balanced risk-benefit information are important, especially in primary prevention.
Place, publisher, year, edition, pages
Blackwell Publishing, 2015. Vol. 26, no 10, p. 1088-1094
Keywords [en]
Complication, hypertrophic cardiomyopathy, implantable defibrillator, inappropriate shock, Sprint Fidelis lead, sudden cardiac death
National Category
Cardiology and Cardiovascular Disease Surgery
Identifiers
URN: urn:nbn:se:oru:diva-122121DOI: 10.1111/jce.12750ISI: 000363211600014PubMedID: 26178879Scopus ID: 2-s2.0-84945195527OAI: oai:DiVA.org:oru-122121DiVA, id: diva2:1979553
Funder
Region Gavleborg2025-06-302025-06-302025-07-08Bibliographically approved