Non-invasive assessment of free steroid hormones: development of a high-throughput LC-MS/MS method for salivary steroid hormone quantificationShow others and affiliations
2025 (English)In: Analytical and Bioanalytical Chemistry, ISSN 1618-2642, E-ISSN 1618-2650, Vol. 417, no 21, p. 4883-4894Article in journal (Refereed) Published
Abstract [en]
Steroid hormone concentrations reflect diverse physiological and pathological processes and have been recognized as valuable biomarkers for disease, with growing interest in their potential for patient stratification in precision medicine. Salivary steroid concentrations should reflect the free (biologically active) steroids in circulation, as steroids in the bloodstream passively diffuse to saliva. This allows for the direct measurement of free steroids without transporter protein-bound hormones (inactive form). However, implementation of salivary steroid quantification in larger studies remains limited by challenges associated with sample preparation. We began by reviewing the literature on relevant sample preparation approaches and identified commonly used 96-well solid phase extraction (SPE) methods for further evaluation. Three sample preparation methods were selected and assessed in terms of internal standard recovery and matrix effects, and their response was also compared using electrospray (ESI) versus UniSpray ionization (USI) liquid chromatography-tandem mass spectrometry (LC-MS/MS). We demonstrate a sensitive and rapid high-throughput method with Oasis HLB mu Elution SPE of 200 mu L saliva in 96-well format and USI-LC-MS/MS for major steroids (testosterone, androstenedione, cortisone, cortisol, and progesterone) in saliva. The method achieved optimal recovery (77%), matrix effects (33%), and sensitivity with detection limits (MDL) ranging between 1.1 and 3.0 pg/mL and linearity (r2 = 0.99). Intra-plate and inter-plate coefficient of variation (CV) was below 7% and 20% using USI. USI provided a higher (2.0-2.8-fold) response than ESI and higher signal-to-noise ratio (S/N). The method was then applied to 97 authentic saliva samples (41 male and 56 female) and significant correlations between age and BMI, and androgen levels were observed in both sexes. The proposed 96-well SPE USI-LC-MS/MS method is well-suited for determining steroid hormones in saliva with potential usefulness in large-scale studies and clinical settings.
Place, publisher, year, edition, pages
Springer, 2025. Vol. 417, no 21, p. 4883-4894
Keywords [en]
Extraction, Free steroid hormones, Saliva, LC-MS/MS, Unispray ionization
National Category
Medical Biotechnology (Focus on Cell Biology, (incl. Stem Cell Biology), Molecular Biology, Microbiology, Biochemistry or Biopharmacy)
Identifiers
URN: urn:nbn:se:oru:diva-122600DOI: 10.1007/s00216-025-06005-9ISI: 001531093800001PubMedID: 40681913Scopus ID: 2-s2.0-105010906858OAI: oai:DiVA.org:oru-122600DiVA, id: diva2:1986546
Funder
Örebro UniversityRegion Östergötland, RÖ-1000278Knowledge Foundation, 20200257
Note
Open access funding provided by Örebro University. Financial support was received from Region Östergötland, Linköping, Sweden (Grant No. RÖ-1000278 to Elisabeth Aardal), and the Swedish Knowledge Foundation (Grant No. 20200257 to Samira Salihovic). Region Östergötland, RÖ-1000278, Stiftelsen för Kunskaps- och Kompetensutveckling, 20200257