Independent thesis Advanced level (degree of Master (Two Years)), 20 credits / 30 HE credits
Background: An airway obstruction occurs when narrowing of the air passages occurs, mainly during expiration, causing an airflow limitation. Asthma and chronic obstructive pulmonary disease (COPD) are two diseases where airflow limitation is present. They are treated pharmacologically with inhalers, mainly dry powder inhalers (DPIs) or pressurised metered-dose inhalers (pMDIs), containing bronchodilators and/or corticosteroids. Airflow limitation is most commonly diagnosed by dynamic spirometry, where Forced Expiratory Volume (in one second) (FEV1) is divided by forced vital capacity (FVC). There is a notion that patients with severe airflow limitation can not use DPIs due to impaired inhalation. Thus, the correlation between FEV1 and peak inhalation flow rate (PIF) needs to be established, and whether the severity of airflow limitation should be used to determine if patients can use DPIs.
Aim: This study aimed to see which lung function parameters, if any, correlate to PIF and if patients with severe airflow limitation can achieve a sufficient PIF for adequate dose-delivery set at 0.8L/s.
Material & Method: 16,000 lung function tests performed over 22 years at two different sites were gathered in a database in Excel.
Results: A weak correlation between FEV1% predicted and PIF could be seen at both sites. A moderate correlation between FVC and PIF was observed at one of the sites. 98% achieved a PIF of 0.8L/s, seen as sufficient inhalation capability.
Conclusion: The severity of obstruction, measured as FEV1 in % predicted, has almost no value when assessing inhalation capacity in individual patients. Even patients with severe airflow limitation achieved a sufficient PIF to reach the required inhalation flow for multi-dose dry powder inhalers. Finally, this study demonstrates that spirometry procedures can vary between sites, and therefore, care should be taken when using data from large databases.
2025.