Optimering av stråldos och bildkvalitet vid konventionell thoraxdatortomografi med tennfilter - En kvantitativ fantomstudie
2026 (Swedish)Independent thesis Basic level (degree of Bachelor), 10 credits / 15 HE credits
Student thesisAlternative title
Optimization of radiation dose and image quality inconventional thoracic computed tomography using tin filter - A quantitative phantom study (English)
Abstract [sv]
Introduktion: Thoraxundersökningar med datortomografi är en vanlig undersökning som medför relativt hög stråldos, jämfört med konventionell slätröntgen. Spektral formning genom tennfilter har potential att minska patientdosen betydligt, särskilt i högkontrastområden som lungor och mediastinum, utan att försämra diagnostisk bildkvalitet.
Syfte: Undersöka hur tennfilter i kombination med varierande Image Quality (IQ) level påverkar stråldos och bildkvalitet vid konventionell thoraxdatortomografi jämfört med standardprotokoll och standardfilter.
Metod: Kvantitativ experimentell fantomstudie, utförd med konventionell datortomografi och ett antropomorfiskt fantom i vuxenstorlek. Ett referensprotokoll med standardfilter jämfördes med sju serier med tennfilter och IQ level mellan 110 och 80. Stråldos och bildkvalitet analyserades subjektivt och objektivt genom Computed Tomography Dose Index by Volume(CTDIvol), Dos Längd Produkt (DLP), signal-brusförhållande (SNR), kontrast-brusförhållande(CNR) och Figure of Merit (FOM).
Resultat: Tennfilterprotokoll reducerade CTDIvol med upp till 39 % vid IQ level 80. I samtliga serier var SNR högre än i referensprotokollet, medan CNR fluktuerade.Doseffektiviteten var som högst vid IQ level 100 i denna studie och den diagnostiska användbarheten av protokollen bedömdes likvärdig genom subjektiv bedömning.
Slutsats: Användning av tennfilter möjliggör en dosreduktion upp till 39% i kombination med förbättrad SNR jämfört med standardprotokoll. CNR varierade mellan olika IQ level, med en tendens till försämring vid lägre IQ level. Trots dessa variationer bedömdes den subjektiva bildkvaliteten likvärdig mellan protokollen av samtliga tre radiologer, vilket indikerar attdiagnostisk bildkvalitet kan bibehållas och ibland förbättras trots reducerad stråldos. Optimal balans mellan stråldos och bildkvalitet definierades vid IQ level 100 med tennfilter.
Abstract [en]
Introduction: Thoracic computed tomography (CT) is a common procedure associated with relatively high radiation dose in comparison with conventional radiography. Tin filtration is a form of spectral shaping with potential to significantly reduce patient dose, especially in high contrastareas such as the lungs and mediastinum, without compromising diagnostic imagequality.
Aim: To establish how tin filtration and variations in Image Quality (IQ) level affect radiation dose and image quality in conventional thoracic CT in comparison with standard filtration.
Method: A quantitative experimental phantom study was conducted using a conventional CT scanner and adult sized chest phantom. A reference protocol with standard filter was compared with seven series using tin filter and IQ levels between 110 and 80. Radiation dose and image quality were analyzed subjectively and objectively with Computed Tomography Dose Index by volume (CTDIvol), Dose Length Product (DLP), Signal-to-Noise-ratio (SNR),Contrast-to-Noise-ratio (CNR) and Figure of Merit (FOM)
Results: Tin filter protocols reduced CTDIvol with up to 39 % at IQ level 80. The CNR fluctuated without linear relationship while the SNR value was higher in all tin filterprotocols. Dose efficiency was highest at IQ level 100 in this study and subjective image quality was assessed comparable across all protocols in terms of diagnostic usability.
Conclusion: The use of tin filter in combination with varying IQ levels allows for a substantial reduction in radiation dose up to 39 % and increased SNR in comparison with the reference protocol, whilst the CNR fluctuated. Despite this variation the subjective image quality was assessed as comparable between all protocols by all participating radiologists,indicating that diagnostic image quality can be maintained and sometimes improved despite reduction in patient dose. An optimal balance between radiation dose and image quality was defined at IQ level 100 using tin filtration.
Place, publisher, year, edition, pages
2026.
Keywords [en]
Computed tomography, tin filter, spectral shaping, image quality, radiation dose
Keywords [no]
Datortomografi, fantom, spektral formning, stråldosoptimering, tennfilter
National Category
Medical and Health Sciences Clinical Medicine Radiology and Medical Imaging
Identifiers
URN: urn:nbn:se:oru:diva-128895OAI: oai:DiVA.org:oru-128895DiVA, id: diva2:2060526
Subject / course
Radiografi
Supervisors
Examiners
2026-05-182026-05-182026-05-18Bibliographically approved